Multielemental analysis using ICP-MS (Type-II)
OIV-MA-AS323-07 Multielemental analysis using ICP-MS
Type II method
- Scope of application
This method can be applied to the analysis of the elements present in wines within the range indicated and featured in the following list:
- Aluminium between 0.25 and 5.0 mg/l
- Boron between 10 and 40 mg/l
- Bromine between 0.20 and 2.5 mg/l
- Cadmium between 0.001 and 0.040 mg/l
- Cobalt between 0.002 and 0.050 mg/l
- Copper between 0.10 and 2.0 mg/l
- Strontium between 0.30 and 1.0 mg/l
- Iron between 0.80 and 5.0 mg/l
- Lithium between 0.010 and 0.050 mg/l
- Magnesium between 50 and 300 mg/l
- Manganese between 0.50 and 1.5 mg/l
- Nickel between 0.010 and 0.20 mg/l
- Lead between 0.010 and 0.20 mg/l
- Rubidium between 0.50 and 1.2 mg/l
- Sodium between 5 and 30 mg/l
- Vanadium between 0.003 and 0.20 mg/l
- Zinc between 0.30 and 1.0 mg/l
This technique can also be used to analyze other elements.
The sample sometimes requires mineralization. This is the case, for example, of wines with more than 100 g/L of sugar where it can be necessary to realise mineralization of the sample before. In this case, it is recommended to perform a digestion with nitric acid in a microwave.
The technique can also be applied to musts, after mineralization.
- Basis
Multielemental quantitative determination using Inductively Coupled Plasma Mass Spectometry or ICP-MS.
Injection and nebulization of the sample in high-frequency plasma. The plasma causes the desolvation, atomization and ionization of the elements in the sample. The ions are extracted using a vacuum system fitted with ionic lenses. The ions are separated according to the mass-to-charge ratio in a mass spectrometer, for example, a quadrupole. Detection and quantification of ions using an electron multiplier system.
- Reagents and solutions
3.1. Ultrapure, demineralized water with resistivity ( 18 MΩ), in accordance with ISO 3696.
3.2. Certified solution(s) (for example, 100 mg/l) containing the metals to be analyzed. Multielemental or monoelemental solutions can be used.
3.3. Indium and/or rhodium solution as an internal standard (normally 1 g/l).
3.4. Nitric acid 60% (metal impurities 0.1 μg/l).
3.5. Argon, minimum purity of 99.999%.
3.6. Nitrogen (maximum impurity content: H2O 3 mg/l, O2 2 mg/l and
CnHm 0.5 mg/l).
Solution concentration and internal standards are given by way of reference.
Preparation of standard solutions:
Acid concentration in the standards and in the final dilution of the wine samples must be the same and must not exceed 5%. The following is an example.
3.7. Stock solution (5mg/l).
Place 0.5ml of solution (3.2) in a 10 ml (4.5) tube and add 0.1 ml of nitric acid (3.4). Level off to 10 ml with demineralized water (3.1) and homogenize.
Shelf life: 1 month.
3.8. Internal standard solution (1 mg/l).
Using micropipettes (4.4), place 50 µl of indium or rhodium solution (3.3) and 0.5 ml of nitric acid (3.4) in a 50 ml tube (4.6). Level off to 50 ml with demineralized water (3.1) and homogenize.
Shelf life: 1 month.
3.9. Standard solutions of the calibration curve.
Adapt the range of the series of standards according to the dilution on the sample or the equipment used.
Use 1000 μl and 100 µl pipettes (4.4).
Shelf life of standard solutions: 1 day
These standard solutions can also be prepared gravimetrically. Add internal standard in the same concentration as for the samples.
3.10. Internal control wine of known concentrations (MRC, MRE, MRI, etc.).
- Material and equipment
4.1. Inductively coupled mass spectrometer with/without collision/reaction cell.
4.2. Computer with data processing software and printer.
4.3. Autosampler (optional).
4.4. 1000 μl and 100 µl micropipettes
4.5. 10 ml plastic, graduated test tubes with bung or glass volumetric flasks.
4.6. 50 ml plastic, graduated test tubes with bung or glass volumetric flasks.
All volumetric material (micropipettes and test tubes) must be duly calibrated.
Note: material that will come into contact with the sample, such as, for example, tubes and tips, must remain for at least 24 hours in a nitric acid solution (3.4) at a concentration of 10% and must subsequently be rinsed several times in water (3.1).
- Sample preparation
Samples of sparkling wine must be degasified. This can be done through nitrogen bubbling (3.6) for 10 minutes or by using an ultrasound bath.
Remove the bung carefully to ensure that the wine is not contaminated. Wash the bottle neck in an acid solution (2% H). Wine samples are taken directly from the bottle.
Use a micropipette (4.4) to insert 0.5 ml of wine, 0.1 to 0.5 ml of nitric acid (3.4) and 100 μl of internal standard solution (3.8) into a 10 ml tube (3.5).
Level off with water (3.1) and homogenize.For certain elements a higher dilution may be necessary owing to their high natural content in the sample.
Br has high ionization potential and its ionization in plasma may be incomplete because of the presence of high concentrations of other elements in wines with low ionization potential. This may result in the incorrect quantification of Br and therefore a 1/50 dilution is recommended to avoid this effect (in the event of another dilution being used, confirm the results by checking recovery after an addition).
When the standards are prepared gravimetrically, the final dilution of the sample must also be obtained gravimetrically.
- Procedure
Switch on the device (pump working and plasma on).
Clean the system for 20 minutes using 2% nitric acid (3.4).
Check that the device is functioning correctly.
Analyze a blank and the series of standard solutions in increasing concentrations, then a standard solution (e.g. no. 2 of series 3.9) to check for correct calibration and finally the blank to ensure that there is no memory effect. Read the samples in duplicate. For the internal control, use a wine of known concentrations (3.10) to confirm that the results are coherent.
Element |
m/z* |
Aluminium |
27 |
Boron |
11 |
Bromine |
79 |
Cadmium |
114 |
Cobalt |
59 |
Copper |
63 |
Strontium |
88 |
Iron |
56/57 |
Lithium |
7 |
Magnesium |
24 |
Manganese |
55 |
Nickel |
60 |
Lead |
average of 206, 207 and 208 |
Rubidium |
85 |
Sodium |
23 |
Vanadium |
51 |
Zinc |
64 |
*The above table is given by way of example. Other isotopes may be required, depending on the equipment.
In the event of using equipment with no collision/reaction cell, correction equations may be necessary for some elements.
- Results
The software can calculate the results directly.
Element concentrations are reported in mg/l to two decimal points.
Obtain, by interpolating in the calibration curve, the concentration of the elements in the diluted samples. Use the following equation to calculate the concentration of the elements in the sample:
|
Where:
C = Concentration of the element in the sample
= Concentration of the elements in the diluted sample
= Final volume of the measurement solution, in ml
= Aliquot volume of wine, in ml.
- Quality control
Ensure traceability by using certified standards.
In each analytical series, use a CRM (Certified Reference Material) as an internal control of wine or a wine used as reference material from an interlaboratory test campaign.
It is recommended that control graphs be created from the results of the quality control analysis.
Participation in interlaboratory test campaigns.
- Precision
The results of the statistical parameters of the collaborative trial are shown in Appendix A.
9.1. Repeatability (r)
The difference between two independent results, obtained using the same method, in the same sample, in the same laboratory, by the same operator, using the same equipment in a short time interval. r results are given in Tables 1 to 17 of Appendix A
9.2. Reproducibility (R)
The difference between two results, obtained using the same method, in the same sample, in a different laboratory, by a different operator and with different equipment. R results are given in Tables 1 to 17 of the Appendix A.
Table 1 represents the % of the relative standard deviation of Repeatability and Reproducibility (RSDr% et RSDR%) of the method. (*) C = Concentration
Element |
Concentration |
RSDr % |
RSDR % |
Aluminium |
0,25 – 5,0 mg/l |
4 |
10 |
Boron |
10 - 40 mg/l |
3,8 |
6,3 |
Bromine |
0,20– 1,0 mg/l |
4,1 |
16,3 |
≥ 1,0 – 2,5 mg/l |
2.1 |
8,0 |
|
Cadmium |
0,001 – 0,020 mg/l |
0,06 C*+0,18 |
10 |
≥ 0,020 – 0,040 mg/l |
1,5 |
10 |
|
Cobalt |
0,002 – 0,050 mg/l |
3,2 |
13,2 |
Copper |
0,10 – 0,50 mg/l |
3,8 |
11,4 |
≥ 0,50 – 2,0 mg/l |
2,0 |
11,4 |
|
Strontium |
0,30 – 1,0 mg/l |
2,5 |
7,5 |
Iron |
0,80– 1,0 mg/l |
4,2 |
15,7 |
≥ 1,0-5,0 mg/l |
4,2 |
7,8 |
|
Lithium |
0,010 – 0,050 mg/l |
7 |
12 |
Magnesium |
50 - 300 mg/l |
2 |
6 |
Manganese |
0,50-1,5 mg/l |
3 |
7 |
Nickel |
0,010 – 0,20 mg/l |
5 |
8 |
Lead |
0,010 – 0,050 mg/l |
8 |
7 |
≥ 0,050 – 0,20 mg/l |
2 |
7 |
|
Rubidium |
0,50 – 1,2 mg/l |
3 |
6 |
Sodium |
5 - 10 mg/l |
2 |
10 |
≥ 10 - 30 mg/l |
0,3 C*-2,5 |
10 |
|
Vanadium |
0,003 – 0,010 mg/l |
8 |
10 |
≥ 0,010 – 0,20 mg/l |
3 |
10 |
|
Zinc |
0,30 – 1,0 mg/l |
5 |
12 |
Table 1: relative standard deviation of Repeatability and Reproducibility
- Bibliography
- ISO 5725:1994, Precision of test methods-Determination of repeatability and reproducibility for a Standard test method by interlaboratory test.
- ISO 17294:2004.
- ALMEIDA M. R, VASCONCELOS T, BARBASTE M. y MEDINA B. (2002), Anal. Bioanal Chem., 374, 314-322.
- CASTIÑEIRA et al. (2001), Frenesius J. Anal. Chem., 370, 553-558.
- DEL MAR CASTIÑEIRA GOMEZ et al. (2004), J. Agric Food Chem., 52, 2962-2974.
- MARISA C., ALMEIDA M. et VASCONCELOS T. (2003), J. Agric. Food Chem., 51, 3012-3023.
- MARISA et al., (2003), J. Agric Food Chem., 51, 4788-4798.
- PÉREZ-JORDAN M. Y., SOLDEVILLA J., SALVADOR A., PASTOR A y de la GURDIA M. (1998), J. Anat. At. Spectrom., 13, 33-39.
- PEREZ-TRUJILLO J.-P., BARBASTE M. y MEDINA B. (2003), Anal. Lett., 36(3), 679-697.
- TAYLOR et al. (2003), J. Agric Food Chem., 51, 856-860.
- THIEL et al. (2004), Anal. Bioanal. Chem, 378, 1630-1636.
Appendix A: Results of the collaborative trials
The method has been checked with two collaborative trials, by evaluating precision in accordance with ISO 5725.The trueness of the method has been obtained through recovery studies.
1st Collaborative Trial
8 samples (A, B, C, D, E, F, MH1 and MH2) were used from the following origins:
- Three samples of red wine, with and without addition.
- Three samples of white wine, with and without addition.
- Two samples of synthetic hydroalcoholic mixture, prepared with ethanol and water.
Hydroalcoholic sample MH1 presented problems of instability during the trial and the results have not been taken into account.
MH2 |
A |
B |
C |
D |
E |
F |
|
Metal (mg/l) |
Hydroalcoholic mixture |
RW2 |
RW3 |
WW2 |
WW3 |
Natural red wine |
Natural white wine |
Aluminium |
5 |
0.5 |
2 |
2 |
1 |
No addition |
No addition |
Cadmium |
0.001 |
0.005 |
0.02 |
0.05 |
0.01 |
No addition |
No addition |
Strontium |
0.300 |
No addition |
No addition |
No addition |
No addition |
No addition |
No addition |
Lithium |
0.020 |
0.01 |
0.02 |
0.04 |
0.01 |
No addition |
No addition |
Magnesium |
50 |
100 |
200 |
50 |
25 |
No addition |
No addition |
Manganese |
0.500 |
0.5 |
1 |
1 |
0.5 |
No addition |
No addition |
Nickel |
0.070 |
0.025 |
0.2 |
0.1 |
0.1 |
No addition |
No addition |
Lead |
0.010 |
0.05 |
0.1 |
0.15 |
0.05 |
No addition |
No addition |
Rubidium |
1.0 |
No addition |
No addition |
No addition |
No addition |
No addition |
No addition |
Sodium |
20 |
10 |
10 |
20 |
5 |
No addition |
No addition |
Vanadium |
0.010 |
0.05 |
0.2 |
0.1 |
0.1 |
No addition |
No addition |
Zinc |
0.500 |
0.1 |
1 |
0.5 |
0.5 |
No addition |
No addition |
2nd Collaborative Trial
Sixteen samples (A, B, C, D, E, F, G, H, I, J, K, L, M, N, O, P) from the following origins were used:
Four samples of red wine, with and without addition.
Four samples of Port wine, with and without addition.
Six samples of white wine, with and without addition.
Two samples of champagne.
Amounts added to the samples
Samples |
Code |
Addition |
B |
Co |
Cu |
Fe |
mg/l |
μg/l |
mg/l |
Mg/l |
|||
White wine |
F-N |
No addition |
0.0 |
0.0 |
0.0 |
0.0 |
C-I |
Addition 1 |
5.0 |
5.0 |
5.0 |
1.0 |
|
A-O |
Addition 2 |
10.0 |
10.0 |
1.0 |
2.0 |
|
Liqueur wine |
B-K |
No addition |
0.0 |
0.0 |
0.0 |
0.0 |
E-L |
Addition 3 |
15.0 |
20.0 |
1.5 |
3.0 |
|
Red wine |
D-M |
No addition |
0.0 |
0.0 |
0.0 |
0.0 |
H-J |
Addition 4 |
20.0 |
50.0 |
2.0 |
5.0 |
|
Sparkling wine |
G-P |
No addition |
0.0 |
0.0 |
0.0 |
0.0 |
Precision parameters (Tables 1 to 17)
The values of Horratr and HorratR have been obtained by using the Horwitz equation taking into account Thompson’s modification for the concentration below 120 µg/L.
Table 1: Aluminium (mg/l)
SAMPLE |
LAB. No. |
Accepted |
Vréf |
Sr |
r |
RSD r (%) |
Horwitz RSDr (%) |
Horratr |
SR |
R |
RSDR (%) |
HorwitzR RSDR (%) |
HorratR |
A |
11 |
10 |
0,68 |
0,020 |
0,06 |
2,9 |
11 |
0,26 |
0,077 |
0,22 |
11 |
17 |
0,66 |
B |
11 |
9 |
2,1 |
0,043 |
0,12 |
2,0 |
9,4 |
0,22 |
0,21 |
0,61 |
10 |
14 |
0,71 |
C |
11 |
9 |
2,1 |
0,032 |
0,09 |
1,5 |
9,5 |
0,16 |
0,21 |
0,59 |
10 |
14 |
0,69 |
D |
11 |
10 |
1,2 |
0,041 |
0,12 |
3,4 |
10 |
0,34 |
0,10 |
0,29 |
8,3 |
16 |
0,56 |
E |
11 |
10 |
0,34 |
0,014 |
0,04 |
4,1 |
12 |
0,34 |
0,029 |
0,08 |
8,5 |
19 |
0,46 |
F |
11 |
10 |
0,27 |
0,006 |
0,02 |
2,2 |
13 |
0,17 |
0,028 |
0,08 |
10 |
20 |
0,52 |
MH2 |
11 |
8 |
5,2 |
0,26 |
0,73 |
5,0 |
8,2 |
0,60 |
0,56 |
1,6 |
11 |
13 |
0,86 |
Table 2: Boron (mg/l)
SAMPLE: |
LAB. No. |
Accepted |
Vréf |
Sr |
r |
RSD r (%) |
Horwitz RSDr (%) |
Horratr |
SR |
R |
RSDR (%) |
HorwitzR RSDR (%) |
HorratR |
A-O |
8 |
6 |
18 |
0,77 |
2,2 |
4,3 |
6,8 |
0,62 |
0,94 |
2,69 |
5,2 |
10 |
0,50 |
B-K |
8 |
4 |
4,5 |
0,27 |
0,76 |
6,0 |
8,4 |
0,72 |
0,40 |
1,14 |
8,9 |
13 |
0,70 |
C-I |
8 |
4 |
13 |
0,31 |
0,89 |
2,4 |
7,2 |
0,33 |
0,33 |
0,94 |
2,5 |
11 |
0,24 |
D-M |
8 |
7 |
11 |
0,26 |
0,74 |
2,4 |
7.4 |
0,31 |
1,1 |
3,11 |
10 |
11 |
0,90 |
E-L |
8 |
5 |
21 |
0,47 |
1,3 |
2,2 |
6.7 |
0,33 |
0,85 |
2,43 |
4,0 |
10 |
0,40 |
F-N |
8 |
5 |
8,3 |
0,43 |
1,2 |
5,2 |
7.7 |
0,68 |
0,47 |
1,34 |
5,7 |
12 |
0,48 |
G-P |
7 |
4 |
3,1 |
0,094 |
0,27 |
3,0 |
8.9 |
0,34 |
0,18 |
0,51 |
5,8 |
14 |
0,43 |
H-J |
8 |
5 |
31 |
1,0 |
3,0 |
3,2 |
6.3 |
0,54 |
1,6 |
4,43 |
5,2 |
9,6 |
0,52 |
Table 3: Bromine (mg/l)
SAMPLE: |
LAB. No. |
Accepted |
Vref |
Sr |
r |
RSD r (%) |
Horwitz RSDr (%) |
Horratr |
SR |
R |
RSDR (%) |
HorwitzR RSDR (%) |
HorratR |
A-O |
6 |
2 |
1,21 |
0,028 |
0,08 |
2,3 |
10,3 |
0,22 |
0,041 |
0,12 |
3,4 |
15,6 |
0,22 |
B-K |
5 |
2 |
0,19 |
0,006 |
0,02 |
2,9 |
13,6 |
0,21 |
0,0043 |
0,012 |
2,3 |
20,5 |
0,11 |
C-I |
6 |
3 |
0,81 |
0,017 |
0,05 |
2,1 |
10,9 |
0,19 |
0,062 |
0,18 |
7,7 |
16,5 |
0,47 |
D-M |
6 |
4 |
0,38 |
0,017 |
0,05 |
4,5 |
12,2 |
0,37 |
0,066 |
0,19 |
17,4 |
18,5 |
0,94 |
E-L |
6 |
3 |
1,72 |
0,030 |
0,09 |
1,7 |
9,7 |
0,17 |
0,22 |
0,62 |
12,8 |
14,8 |
0,86 |
F-N |
6 |
3 |
0,22 |
0,014 |
0,04 |
6,4 |
13,3 |
0,48 |
0,046 |
0,13 |
20,9 |
20,1 |
1 |
H-J |
6 |
2 |
2,30 |
0,061 |
0,17 |
2.7 |
9,3 |
0,28 |
0,092 |
0,26 |
4 |
14.1 |
0.28 |
Table 4: Cadmium (μg/l)
SAMPLE: |
LAB. No. |
Accepted |
Vréf |
Sr |
r |
RSD r (%) |
Horwitz RSDr (%) |
Horratr |
SR |
R |
RSDR (%) |
HorwitzR RSDR (%) |
HorratR |
A |
12 |
11 |
6 |
0,2 |
0,6 |
3,3 |
15 |
0,22 |
1 |
3 |
17 |
22 |
0,77 |
B |
12 |
11 |
16 |
0,4 |
1 |
2,5 |
15 |
0,17 |
2 |
6 |
13 |
22 |
0,59 |
C |
12 |
9 |
40 |
0,4 |
1 |
1,0 |
15 |
0,07 |
3 |
8 |
7,5 |
22 |
0,34 |
D |
12 |
10 |
10 |
0,3 |
0,8 |
3,0 |
15 |
0,20 |
0,9 |
3 |
9,0 |
22 |
0,41 |
E |
8 |
7 |
0,3 |
0,20 |
0,6 |
67 |
15 |
4,47 |
0,20 |
0,67 |
67 |
22 |
3,05 |
F |
8 |
6 |
0,3 |
0,04 |
0,1 |
13 |
15 |
0,87 |
0,20 |
0,45 |
67 |
22 |
3,05 |
MH2 |
9 |
5 |
0,9 |
0,08 |
0,2 |
8,9 |
15 |
0,59 |
0,10 |
0,29 |
11 |
22 |
0,50 |
Table 5: Cobalt (μg/l)
SAMPLE: |
LAB. No. |
Accepted |
Vréf |
Sr |
r |
RSD r (%) |
Horwitz RSDr (%) |
Horratr |
SR |
R |
RSDR (%) |
HorwitzR RSDR (%) |
HorratR |
A-O |
10 |
6 |
22 |
0,5 |
1 |
2,3 |
15 |
0,15 |
2 |
6 |
9,1 |
22 |
0,41 |
B-K |
10 |
6 |
8 |
0,3 |
0,9 |
3,8 |
15 |
0,25 |
1 |
4 |
13 |
22 |
0,59 |
C-I |
10 |
8 |
19 |
0,4 |
1 |
2,1 |
15 |
0,14 |
3 |
7 |
16 |
22 |
0,73 |
D-M |
10 |
3 |
3 |
0,07 |
0,2 |
2,3 |
15 |
0,15 |
0,1 |
0,3 |
3,3 |
22 |
0,15 |
E-L |
10 |
8 |
27 |
1 |
3 |
3,7 |
15 |
0,25 |
3 |
9 |
11 |
22 |
0,50 |
F-N |
10 |
7 |
12 |
0,5 |
2 |
4,2 |
15 |
0,28 |
1 |
4 |
8,3 |
22 |
0,38 |
G-P |
9 |
5 |
2 |
0,2 |
0,5 |
10 |
15 |
0,67 |
0,3 |
0,8 |
15 |
22 |
0,68 |
H-J |
10 |
6 |
49 |
0,5 |
1 |
2,3 |
15 |
0,15 |
6 |
18 |
12 |
22 |
0,55 |
Table 6: Copper (mg/l)
SAMPLE |
LAB. No. |
Accepted |
Vréf |
Sr |
r |
RSD r (%) |
Horwitz RSDr (%) |
Horratr |
SR |
R |
RSDR (%) |
HorwitzR RSDR (%) |
HorratR |
A-O |
10 |
8 |
1,1 |
0,013 |
0,040 |
1,2 |
10 |
0,12 |
0,11 |
0,32 |
10 |
16 |
0,63 |
B-K |
10 |
8 |
0,21 |
0,006 |
0,020 |
2,9 |
13 |
0,22 |
0,021 |
0,060 |
10 |
20 |
0,50 |
C-I |
10 |
7 |
0,74 |
0,009 |
0,030 |
1,2 |
10 |
0,12 |
0,046 |
0,13 |
6,2 |
17 |
0,36 |
D-M |
10 |
8 |
0,14 |
0,007 |
0,020 |
5,0 |
14 |
0,36 |
0,015 |
0,043 |
11 |
22 |
0,50 |
E-L |
10 |
9 |
1,7 |
0,061 |
0,17 |
3,6 |
7,8 |
0,5 |
0,16 |
0,46 |
9,0 |
15 |
0,60 |
F-N |
10 |
7 |
0,16 |
0,006 |
0,020 |
3,8 |
14 |
0,27 |
0,029 |
0,083 |
18 |
21 |
0,86 |
G-P |
9 |
4 |
0,042 |
0,004 |
0,010 |
9,5 |
15 |
0,63 |
0,006 |
0,017 |
14 |
22 |
0,64 |
H-J |
10 |
7 |
2,1 |
0,018 |
0,050 |
0,86 |
9,5 |
0,09 |
0,24 |
0,69 |
11 |
14 |
0,79 |
Table 7:
Strontium (μg/l)
SAMPLE |
LAB. Nº |
Accepted |
Vréf |
Sr |
r |
RSD r (%) |
Horwitz RSDr (%) |
Horratr |
SR |
R |
RSDR (%) |
HorwitzR RSDR (%) |
HorratR |
|
A |
12 |
11 |
1091 |
33 |
93 |
3,0 |
10 |
0,30 |
78 |
222 |
7,2 |
16 |
0,45 |
|
B |
12 |
8 |
1139 |
66 |
188 |
5,8 |
10 |
0,58 |
69 |
195 |
6,1 |
16 |
0,38 |
|
C |
12 |
9 |
328 |
6 |
18 |
1,8 |
13 |
0,14 |
19 |
54 |
5,8 |
19 |
0,31 |
|
D |
12 |
10 |
313 |
7 |
20 |
2,2 |
13 |
0,17 |
22 |
61 |
7,0 |
19 |
0,37 |
|
E |
12 |
10 |
1176 |
28 |
80 |
2,4 |
10 |
0,24 |
86 |
243 |
7,3 |
16 |
0,46 |
|
F |
12 |
10 |
293 |
3 |
9 |
1,0 |
13 |
0,08 |
22 |
62 |
7,5 |
19 |
0,39 |
|
MH2 |
12 |
9 |
352 |
7 |
19 |
2,0 |
12 |
0,17 |
24 |
69 |
6,8 |
19 |
0,36 |
|
Table 8: Iron (mg/l)
SAMPLE |
LAB. No. |
Accepted |
Vréf |
Sr |
r |
RSD r (%) |
Horwitz RSDr (%) |
Horratr |
SR |
R |
RSDR (%) |
HorwitzR RSDR (%) |
HorratR |
A-O |
10 |
6 |
3,2 |
0,017 |
0,05 |
0,53 |
8,9 |
0,06 |
0,23 |
0,66 |
7,2 |
13 |
0,55 |
B-K |
10 |
6 |
1,5 |
0,085 |
0,24 |
5,7 |
9,9 |
0,58 |
0,11 |
0,31 |
7,3 |
15 |
0,49 |
C-I |
10 |
5 |
2,1 |
0,036 |
0,10 |
1,7 |
9,4 |
0,18 |
0,18 |
0,51 |
8,6 |
14 |
0,61 |
D-M |
10 |
5 |
3,1 |
0,033 |
0,094 |
1,1 |
8,9 |
0,12 |
0,29 |
0,83 |
9,4 |
14 |
0,67 |
E-L |
10 |
5 |
4,3 |
0,120 |
0,34 |
2,8 |
8,5 |
0,33 |
0,29 |
0,83 |
6,7 |
13 |
0,52 |
F-N |
10 |
6 |
1,1 |
0,051 |
0,15 |
4,6 |
10 |
0,46 |
0,16 |
0,46 |
15 |
16 |
0,94 |
G-P |
9 |
6 |
0,83 |
0,024 |
0,07 |
2,9 |
11 |
0,26 |
0,14 |
0,40 |
17 |
16 |
1,06 |
H-J |
10 |
7 |
7,8 |
0,180 |
0,52 |
2,3 |
7,8 |
0,29 |
1,2 |
3,52 |
15 |
12 |
1,25 |
Table 9: Lithium (μg/l)
SAMPLE |
LAB. No. |
Accepted |
Vréf |
Sr |
r |
RSD r (%) |
Horwitz RSDr (%) |
Horratr |
SR |
R |
RSDR (%) |
HorwitzR RSDR (%) |
HorratR |
A |
11 |
10 |
34 |
2 |
5 |
5,9 |
15 |
0,39 |
4 |
11 |
11 |
22 |
0,50 |
B |
11 |
11 |
42 |
3 |
8 |
7,1 |
15 |
0,47 |
4 |
12 |
10 |
22 |
0,45 |
C |
11 |
11 |
47 |
1 |
4 |
2,1 |
15 |
0,14 |
5 |
13 |
9,8 |
22 |
0,45 |
D |
11 |
11 |
18 |
1 |
4 |
5,6 |
15 |
0,37 |
2 |
7 |
14 |
22 |
0,64 |
E |
11 |
11 |
25 |
1 |
3 |
4,0 |
15 |
0,27 |
3 |
9 |
12 |
22 |
0,55 |
F |
11 |
9 |
9 |
0,3 |
1 |
3,8 |
15 |
0,25 |
0,6 |
2 |
7,2 |
22 |
0,33 |
MH2 |
11 |
7 |
22 |
1 |
3 |
4,6 |
15 |
0,31 |
1 |
3 |
5,3 |
22 |
0,24 |
Table 10: Magnesium (mg/l)
SAMPLE: |
LAB. No. |
Accepted |
Vréf |
Sr |
r |
RSD r (%) |
Horwitz RSDr (%) |
Horratr |
SR |
R |
RSDR (%) |
HorwitzR RSDR (%) |
HorratR |
A |
10 |
7 |
182 |
2,9 |
8,1 |
1,6 |
4,3 |
0,37 |
9,3 |
26 |
5,1 |
7,3 |
0,70 |
B |
10 |
6 |
280 |
3,9 |
11 |
1,4 |
4,5 |
0,31 |
6,0 |
17 |
2,1 |
6,9 |
0,30 |
C |
10 |
7 |
104 |
2,4 |
6,9 |
2,3 |
5,3 |
0,43 |
6,8 |
19,25 |
6,5 |
8,0 |
0,81 |
D |
10 |
6 |
85 |
1,4 |
4,0 |
1,7 |
5,4 |
0,31 |
2,2 |
6,1 |
2,6 |
8,2 |
0,32 |
E |
10 |
7 |
94 |
2,2 |
6,2 |
2,3 |
5,3 |
0,43 |
5,5 |
16 |
5,9 |
8,1 |
0,73 |
F |
10 |
7 |
65 |
0,95 |
2,7 |
1,5 |
5,6 |
0,27 |
3,8 |
11 |
5,9 |
8,5 |
0,69 |
MH2 |
10 |
7 |
51 |
0,90 |
2,5 |
1,8 |
5,8 |
0,31 |
2,4 |
6,9 |
4,7 |
8,9 |
0,53 |
Table 11: Manganese (mg/l)
SAMPLE |
LAB. No. |
Accepted |
Vréf |
Sr |
r |
RSD r (%) |
Horwitz RSDr (%) |
Horratr |
SR |
R |
RSDR (%) |
HorwitzR RSDR (%) |
HorratR |
A |
11 |
10 |
1,3 |
0,014 |
0,040 |
1,1 |
10 |
0,11 |
0,13 |
0,37 |
10 |
15 |
0,67 |
B |
11 |
9 |
1,8 |
0,14 |
0,40 |
7,8 |
9,7 |
0,80 |
0,20 |
0,56 |
11 |
15 |
0,73 |
C |
11 |
8 |
1,5 |
0,028 |
0,080 |
1,9 |
9,9 |
0,19 |
0,084 |
0,24 |
5,6 |
15 |
0,37 |
D |
11 |
8 |
1,0 |
0,035 |
0,10 |
3,5 |
11 |
0,32 |
0,049 |
0,14 |
4,9 |
16 |
0,31 |
E |
11 |
9 |
0,84 |
0,019 |
0,050 |
2,3 |
11 |
0,21 |
0,057 |
0,16 |
6,8 |
16 |
0,43 |
F |
11 |
9 |
0,59 |
0,015 |
0,040 |
2,5 |
11 |
0,23 |
0,031 |
0,090 |
5,3 |
17 |
0,31 |
MH2 |
11 |
8 |
0,52 |
0,029 |
0,080 |
5,6 |
12 |
0,47 |
0,037 |
0,10 |
7,1 |
18 |
0,39 |
Table 12: Nickel (μg/l)
SAMPLE |
LAB. No. |
Accepted |
Vréf |
Sr |
r |
RSD r (%) |
Horwitz RSDr (%) |
Horratr |
SR |
R |
RSDR (%) |
HorwitzR RSDR (%) |
HorratR |
A |
11 |
10 |
40 |
2 |
6 |
5,0 |
15 |
0,33 |
5 |
13,90 |
13 |
22 |
0,59 |
B |
12 |
10 |
194 |
7 |
20 |
3,6 |
14 |
0,26 |
17 |
48,96 |
8,8 |
21 |
0,42 |
C |
12 |
8 |
148 |
4 |
10 |
2,7 |
14 |
0,19 |
5 |
15,12 |
3,4 |
21 |
0,16 |
D |
12 |
8 |
157 |
4 |
12 |
2,6 |
14 |
0,19 |
8 |
23,10 |
5,1 |
21 |
0,24 |
E |
11 |
8 |
15 |
0,6 |
2 |
4,0 |
15 |
0,27 |
1 |
3,33 |
6,7 |
22 |
0,30 |
F |
12 |
9 |
66 |
1 |
4 |
1,5 |
15 |
0,10 |
4 |
10,58 |
6,1 |
22 |
0,28 |
MH2 |
11 |
7 |
71 |
5 |
14 |
7,0 |
15 |
0,47 |
4 |
11,41 |
5,6 |
22 |
0,25 |
Table 13: Lead (μg/l)
SAMPLE |
LAB. No. |
Accepted |
Vréf |
Sr |
r |
RSD r (%) |
Horwitz RSDr (%) |
Horratr |
SR |
R |
RSDR (%) |
HorwitzR RSDR (%) |
HorratR |
A |
12 |
9 |
59 |
1 |
4 |
1,7 |
15 |
0,11 |
3 |
9 |
5,1 |
22 |
0,23 |
B |
12 |
10 |
109 |
2 |
6 |
1,8 |
15 |
0,12 |
8 |
23 |
7,3 |
22 |
0,33 |
C |
12 |
9 |
136 |
3 |
9 |
2,2 |
14 |
0,16 |
13 |
37 |
9,6 |
22 |
0,44 |
D |
12 |
9 |
119 |
2 |
6 |
1,7 |
15 |
0,11 |
5 |
13 |
4,2 |
22 |
0,19 |
E |
12 |
10 |
13 |
1 |
3 |
7,7 |
15 |
0,51 |
1 |
4 |
7,7 |
22 |
0,35 |
F |
12 |
9 |
92 |
1 |
4 |
1,1 |
15 |
0,07 |
4 |
11 |
4,4 |
22 |
0,20 |
MH2 |
12 |
10 |
13 |
1 |
3 |
7,7 |
15 |
0,51 |
1 |
3 |
7,7 |
22 |
0,35 |
Table 14: Rubidium (μg/l)
SAMPLE |
LAB. No. |
Accepted |
Vréf |
Sr |
r |
RSD r (%) |
Horwitz RSDr (%) |
Horratr |
SR |
R |
RSDR (%) |
HorwitzR RSDR (%) |
HorratR |
A |
11 |
6 |
717 |
14 |
41 |
2,0 |
11 |
0,18 |
13 |
36 |
1,8 |
17 |
0,11 |
B |
11 |
7 |
799 |
25 |
70 |
3,1 |
11 |
0,28 |
30 |
86 |
3,8 |
17 |
0,22 |
C |
11 |
8 |
677 |
10 |
27 |
1,5 |
11 |
0,14 |
34 |
96 |
5,0 |
17 |
0,29 |
D |
11 |
7 |
612 |
18 |
51 |
2,9 |
11 |
0,26 |
18 |
50 |
2,9 |
17 |
0,17 |
E |
11 |
9 |
741 |
19 |
53 |
2,6 |
11 |
0,24 |
66 |
187 |
8,9 |
17 |
0,52 |
F |
11 |
9 |
617 |
10 |
28 |
1,6 |
11 |
0,15 |
43 |
123 |
7,0 |
17 |
0,41 |
MH2 |
11 |
7 |
1128 |
10 |
28 |
0,89 |
10 |
0,09 |
64 |
181 |
5,7 |
16 |
0,36 |
Table 15: Sodium (mg/l)
SAMPLE |
LAB. No. |
Accepted |
Vréf |
Sr |
r |
RSD r (%) |
Horwitz RSDr (%) |
Horratr |
SR |
R |
RSDR (%) |
HorwitzR RSDR (%) |
HorratR |
A |
10 |
9 |
19 |
0,59 |
1,7 |
3,1 |
6,8 |
0,46 |
2,2 |
5,7 |
12 |
10 |
1,20 |
B |
10 |
9 |
20 |
1,3 |
3,6 |
6,5 |
6,7 |
0,97 |
2,2 |
6,3 |
11 |
10 |
1,10 |
C |
10 |
7 |
28 |
0,33 |
0,93 |
1,2 |
6,4 |
0,19 |
1,9 |
5,4 |
6,8 |
9,7 |
0,70 |
D |
10 |
8 |
11 |
0,24 |
0,68 |
2,2 |
7,4 |
0,30 |
1,1 |
3,0 |
10 |
11 |
0,91 |
E |
10 |
8 |
9,8 |
0,19 |
0,53 |
1,9 |
7,5 |
0,25 |
0,89 |
2,5 |
9,1 |
11 |
0,83 |
F |
10 |
8 |
6,1 |
0,093 |
0,26 |
1,5 |
8,1 |
0,19 |
0,74 |
2,1 |
12 |
12 |
1,00 |
MH2 |
10 |
8 |
24 |
1,8 |
5,0 |
7,5 |
6,6 |
1,14 |
2,6 |
7,2 |
11 |
9,9 |
1,11 |
Table 16: Vanadium (μg/l)
SAMPLE |
LAB. No. |
Accepted |
Vréf |
Sr |
r |
RSD r (%) |
Horwitz RSDr (%) |
Horratr |
SR |
R |
RSDR (%) |
HorwitzR RSDR (%) |
HorratR |
A |
12 |
11 |
46 |
1 |
3 |
2,2 |
15 |
0,15 |
5 |
13 |
11 |
22 |
0,50 |
B |
12 |
11 |
167 |
5 |
15 |
3,0 |
14 |
0,21 |
19 |
54 |
11 |
21 |
0,52 |
C |
12 |
11 |
93 |
3 |
8 |
3,2 |
15 |
0,21 |
12 |
33 |
13 |
22 |
0,59 |
D |
12 |
9 |
96 |
3 |
8 |
3,1 |
15 |
0,21 |
8 |
22 |
8,3 |
22 |
0,38 |
E |
10 |
7 |
3 |
0,2 |
0,7 |
6,7 |
15 |
0,45 |
0,3 |
0,9 |
10 |
22 |
0,45 |
F |
10 |
8 |
3 |
0,2 |
0,6 |
6,7 |
15 |
0,45 |
0,2 |
0,7 |
6,7 |
22 |
0,30 |
MH2 |
12 |
9 |
11 |
0,3 |
1 |
2,7 |
15 |
0,18 |
0,9 |
3 |
8,2 |
22 |
0,37 |
Table 17: Zinc (μg/l)
SAMPLE |
LAB. No. |
Accepted |
Vréf |
Sr |
r |
RSD r (%) |
Horwitz RSDr (%) |
Horratr |
SR |
R |
RSDR (%) |
HorwitzR RSDR (%) |
HorratR |
A |
11 |
8 |
405 |
22 |
61 |
5,4 |
12 |
0,45 |
45 |
128 |
11 |
18 |
0,61 |
B |
11 |
9 |
1327 |
49 |
138 |
3,7 |
10 |
0,37 |
152 |
429 |
11 |
15 |
0,73 |
C |
11 |
9 |
990 |
14 |
41 |
1,4 |
11 |
0,13 |
86 |
243 |
8,7 |
16 |
0,54 |
D |
11 |
9 |
1002 |
28 |
79 |
2,8 |
11 |
0,25 |
110 |
310 |
11 |
16 |
0,69 |
E |
11 |
9 |
328 |
13 |
37 |
4,0 |
13 |
0,31 |
79 |
224 |
24 |
19 |
1,26 |
F |
11 |
9 |
539 |
15 |
42 |
2,8 |
12 |
0,23 |
61 |
172 |
11 |
18 |
0,61 |
MH2 |
11 |
8 |
604 |
72 |
204 |
12 |
11 |
1,09 |
89 |
251 |
15 |
17 |
0,88 |