Compendium of International Methods of Wine and Must Analysis

Download document

Alcoholic strength by volume (Type-I-and-IV)

OIV-MA-AS312-01 Alcoholic strength by volume

Type I and type IV methods

 

  1. Introduction

This resolution is made up of one part on distillate preparation, followed by 4 methods (A, B, C and D) to determine the alcoholic strength by volume of this distillate.

  1. Scope of application

 

This resolution is applicable for determining the alcoholic strength by volume at 20 °C of vitivinicultural beverages, using any of the following:

Method A: Pycnometry

Type I Method,

Method B: Electronic densimetry using a frequency oscillator

Type I Method,

Method C: Hydrostatic balance

Type I Method,

Method D: Hydrometry and refractometry

Common Type IV Method.

  1. Definition

The alcoholic strength by volume (ABV) of a beverage is the number of litres of ethanol contained in 100 litres of hydroalcoholic solution with the same density as the beverage distillate; both volumes being determined at a temperature of 20 °C. It is expressed by the symbol '% vol.'.

  1. Principle and methods

 

4.1.  Principle

The principle of the method consists firstly of distilling the beverage by volume to volume after alkalinisation by a suspension of calcium hydroxide, which

prevents the entrainment of volatile acids. This distillation enables the elimination of non-volatile substances. The homologues of ethanol, in addition to ethanol and its homologues in esters are included in the ABV since they are present in the distillate.

Secondly, the density of the distillate is measured. The density of a liquid at a given temperature is equal to the quotient of its mass over its volume:

= m / V, and for a vitivinicultural beverage, it is expressed as g/cm3.

For hydro-alcoholic solutions such as distillates, when the temperature is known, the tables can be used to match the density up to the ABV (OIV, MA-AS312-02: R2009 Table 1). This ABV corresponds to that of the beverage (distillation by volume to volume).

4.2.  Methods of determination of ABV

The principle and procedure for each method are detailed in the following parts:

 Part A: Determination of the alcoholic strength by volume of a beverage by measuring of the density of the distillate using a pycnometer;

Part B: Determination of the alcoholic strength by volume of a beverage by measuring the density of the distillate by electronic densimetry using a frequency oscillator;

Part C: Determination of the alcoholic strength by volume of a beverage by measuring the density of the distillate by densimetry using a hydrostatic balance;

Part D: Determination of the alcoholic strength by volume of a beverage by measuring the density of the distillate by hydrometry or by refractometry.

The test temperature is set at 20 °C.

4.3.  Safety precautions

Respect the safety guidelines for the usage of distillation apparatus, and for the handling of hydro-alcoholic and cleaning solutions.

  1. Obtaining the distillate

5.1.  Reagents

5.1.1.      Type II water for analytical usage (ISO 3696 standard), or of equivalent purity

5.1.2.      Suspension of calcium hydroxide, 12% m/v

Obtain by carefully pouring 1 L of water at 60-70 °C onto 120 g of quicklime (CaO).

5.1.3.      Anti-foaming agent

5.2.  Apparatus

Any type of distillation or steam distillation apparatus may be used provided that it satisfies the following test:

Distil a hydro-alcoholic mixture with an alcoholic strength of 10% vol. five times in succession. The distillate should have an alcoholic strength of at least 9.9% vol. after the fifth distillation, i.e. the loss of alcohol during each distillation should not be more than 0.02% vol.

By way of example, use one of the following two sets of apparatus.

5.2.1.      Distillation apparatus, consisting of:

  • a round bottomed 1-L flask with a ground-glass standard joint,
  • a rectifying column of about 20 cm in height or any system designed to prevent priming,
  • a source of heat (any pyrolysis of extracted matter should be prevented by a suitable arrangement),
  • a condenser terminated by a tapered tube taking the distillate to the bottom of a graduated receiver flask containing several mL of water. 
    1.       Steam distillation apparatus consisting of:
  • a steam generator,
  • a bubbler,
  • a rectifying column,
  • a condenser.
    1.   Preparation of the sample

Remove the bulk of any carbon dioxide from samples with bubbles (e.g. by stirring 250 to 300 mL of the wine in a 500-mL flask).

5.4.  Procedure

5.4.1.      Procedure for beverages with an ABV greater than or equal to 1.5% vol.

Take a sample of a 200-mL volume of beverage using a calibrated flask. Note the temperature of the sample.

Pour it into the flask of the distillation apparatus or into the bubbler of the steam distillation apparatus. Rinse the calibrated flask four times with approx. 5 mL of water and add this to the apparatus' flask or bubbler.

Add approx. 10 mL 2 M calcium hydroxide ((5.1.2). If necessary, several fragments of inert porous material (e.g. pumice, etc.) and/or several drops of anti-foaming agent (5.1.3) may also be added to facilitate distillation.

Collect the distillate in the 200-mL calibrated flask used to measure the beverage. Collect a volume of about three-quarters of the initial volume if distillation is used or 198-199 mL of distillate if steam distillation is used.

Make up to 200 mL with distilled water, keeping the distillate at within ± 2 °C of the initial temperature.

Carefully mix using a circular motion.

Note: In the case of wines containing particularly large concentrations of ammonium ions, the distillate may be redistilled under the conditions described above, but replacing the suspension of calcium hydroxide with 1 mL sulphuric acid diluted to 10% (v/v).

5.4.2.      Procedure for beverages with an ABV less than 1.5% vol.

Take a sample of a 200-mL volume of beverage using a calibrated flask. Note the temperature of the beverage. Pour it into the flask of the distillation apparatus or into the bubbler of the steam distillation apparatus. Rinse the calibrated flask four times with approx. 5 mL of water and add this to the apparatus' flask or bubbler. Add approx. 10 mL 2 M calcium hydroxide (5.1.2) and, in the case of distillation, if necessary, a boiling regulating agent (e.g. pumice, etc.). In a 100-mL calibrated flask, collect a volume of distillate of about 75 mL if distillation is used or 98-99 mL of distillate if steam distillation is used.

Make up to 100 mL with distilled water, keeping the distillate at within 2°C of the initial temperature. Carefully mix using a circular motion.

 

Part A: Determination of the alcoholic strength by volume of a beverage by measuring the density of the distillate using a pycnometer

(Type I Method)

(Method A2/1978 – Resolution OIV/OENO 377/2009)

A.1.           Principle

 

The density of the distillate is determined, which is matched to the ABV using the Tables.

The density is measured for a specific temperature using a glass pycnometer. This comprises a flask of known capacity, onto which a hollow ground-glass stopper is fitted equipped with a capillary tube. When the flask is closed, the overflow rises in the capillary. The volumes of the flask and the capillary being known, the density is determined by weighing using precision balances before and after filling of the pycnometer.

A.2.           Reagents and products

 

A.2.1.    Type II water for analytical usage (ISO 3696 standard), or of equivalent purity

A.2.2.    Sodium chloride solution (2% m/v)

To prepare 1 litre, weigh out 20 g sodium chloride and dissolve to volume with water.

A.3.           Apparatus and material

 

Common laboratory apparatus, including the following:

A.3.1.     Pyrex-glass pycnometer of around 100 mL capacity with a removable thermometer, with ground-glass joint and 10th-of-a-degree graduations, from 10°C to 30°C. This thermometer should be calibrated (Fig. 1).

Any pycnometer of equivalent characteristics may be used.

 

This pycnometer includes a side tube of 25 mm in length and an inside diameter of at most 1 mm, terminated by a ground-glass conical joint. This side tube may be capped by a 'reservoir stopper' composed of a ground-glass conical tube, terminated by a tapered joint. This stopper serves as an expansion chamber.

The two joints of the apparatus should be prepared with great care.

A.3.2.     Tare bottle of the same external volume (to within 1 mL) as the pycnometer and with a mass equal to the mass of the pycnometer filled with a liquid of a density of 1.01 (2% m/v sodium chloride solution).

A.3.3.     Thermally-insulated jacket that fits the body of the pycnometer exactly.

A.3.4.     Twin-pan balance accurate to the nearest 0.1 mg

 or

single-plate balance accurate to the nearest 0.1 mg.

A.3.5.     Masses calibrated by an accredited body.

 

A.4.           Procedure

A.4.1.    Pycnometer calibration

The calibration of the pycnometer comprises the determination of the following characteristics:

  • tare weight,
  • volume at 20 °C,
  • water mass at 20 °C.
    1.                        Using a twin-pan balance

Place the tare bottle on the left-hand pan and the clean, dry pycnometer with its 'reservoir stopper' on the right-hand pan. Balance them by placing weights of known mass on the pycnometer side: p grams.

Fill the pycnometer carefully with water (A.2.1) at room temperature and fit the thermometer.

Carefully wipe the pycnometer dry and place it in the thermally-insulated jacket.

Shake by inverting the container until the thermometer's temperature reading is constant. Accurately adjust the level to the upper rim of the side tube. Wipe the side tube clean and fit the reservoir stopper.

Read the temperature, t °C, carefully and if necessary correct for any inaccuracies in the temperature scale.

Weigh the water-filled pycnometer, with the weight in grams, p, making up the equilibrium.

Calculations

Tare of the empty pycnometer:

Tare weight = p + m where m = mass of the air contained in the pycnometer, in g

  • m (g) = 0.0012 (p - p’)

Volume at 20 °C in mL:

(mL) = (p + m p') x Ft

Ft = factor for temperature, t °C, taken from Table I

should be known to 0.001 mL

Water mass at 20 °C:

= x 0.998203, in g

0.998203 = water density at 20 °C, in g/cm3 

A.4.1.2.        Using a single-pan balance

Determine:

  • the mass of the clean, dry pycnometer: P,
  • the mass of the water-filled pycnometer at t °C: P1 following the instructions outlined in A.4.1.1,
  • the mass of the tare bottle, T0.

Calculations

Tare of the empty pycnometer:

Tare weight: P – m where m (g) = mass of the air contained in the pycnometer, in g

m (g) = 0.0012 ( - P)

Volume at 20 °C in mL:

= factor for temperature, t °C, taken from Table I

should be known to 0.001 mL.

Water mass at 20 °C:

0.998203 = water density at 20 °C, in g/cm3

A.4.2.     Determination of the density of the distillate

Measure the apparent density of the distillate at t °C using a twin-pan or single-pan balance:

A.4.2.1.                       Using a twin-pan balance

Weigh the pycnometer filled with the test sample following the instructions outlined in A.4.1.1.

Where p" represents the mass in grams that makes up the equilibrium at  °C,

taking into account that the liquid mass contained in the pycnometer = p + m - p",

the apparent density at t °C, in g/cm3, is given by the following equation

 

A.4.2.2.                       Using a single-pan balance

Weigh the tare bottle, where T1 is its mass in g.

Calculate

Mass of the empty pycnometer at the time of measurement = P - m + dT, in g

Weigh the pycnometer filled with the test sample following the instructions outlined in A.4.1.1.

Where P2 represents its mass at t °C,

the liquid mass contained in the pycnometer at t °C = (P - m + dT), in g

and the apparent density at t °C, in g/cm3, is as follows:

A.5.           Expression of results and precision parameters

 

A.5.1.     Method of calculation

A.5.1.1.                       Beverages with an ABV greater than or equal to 1.5% vol.

Find the alcoholic strength at 20 °C in % vol. to 2 d.p. using Table I of Method OIV-MA-AS312-02A. Please note, this table uses the unit kg/m3 and not g/cm3.

The relationship is as follows: 1 g/cm3 = 1000 kg/m3.

In the horizontal line of this table corresponding to the temperature, T, (expressed as a whole number) immediately below t °C, find the smallest density greater than t. Use the tabular difference just below this density to calculate the density ρ at this temperature, T.

On the line of the temperature, T, find the density ' immediately above  and calculate the difference between the densities ρ and ρ'. Divide this difference by the tabular difference just to the right of the density ρ'. The quotient gives the decimal part of the alcoholic strength, while the whole number part of this strength is shown at the head of the column in which the density ρ' is located.

An example of calculation of the alcoholic strength is given in Annex 1 to this Chapter.

Note: This temperature correction has been incorporated into a computer program and might possibly be carried out automatically.

A.5.1.2.                       Beverages with an ABV less than 1.5% vol.

Identical to A.5.1.1, dividing the alcoholic strength by volume of the distillate (ABVD) by 2.

ABV = ABVD/2, % vol. to 2 d.p.

A.6.           Precision

 

Repeatability (r):

  • r = 0.10 % vol.

Reproducibility (R)

  • R = 0.19 % vol.

The validation parameters for beverages with a low alcohol content are given in Annex II.

A.7.           Example of the calculation of the alcoholic strength of a wine

A.7.1.    Measurement by pycnometer on a twin-pan balance

The constants of the pycnometer have been determined and calculated as described in the method OIV-MA-AS2-01, ‘Density and specific gravity’, paragraph A.7.

A.4.2.1 Using a twin-pan balance

Numerical example

1. Weighing of the distillate-filled pycnometer:

Tare = pycnometer + distillate at t °C + p”

 

p + mp” = mass of the distillate at t °C

Apparent density at t °C:

2. Calculation of the alcoholic strength:

Consult the table of apparent densities of hydro-alcoholic mixtures at different temperatures, as indicated above.

t °C = 18.90 °C
corrected t °C = 18.70 °C

p”  = 2.8074 g

105.0698 – 2.8074 = 102.2624 g

On the line 18 °C of the table of apparent densities, the smallest density greater than the observed density of 0.983076 is 0.98398, in the column 11%.

The density at 18 °C is:

(98307.6 + 0.7 x 22) 10-5 = 0.98323  0.98398 – 0.98323 = 0.00075

The decimal portion of the alcoholic strength is: 75 / 114 = 0.65

The alcoholic strength is: 11.65% vol.

A.7.2.     Measurement by pycnometer on a single-pan balance

The constants of the pycnometer have been determined and calculated as described in the method OIV-MA-AS2-01, ‘Density and specific gravity’, paragraph A.7.

Numerical example

Weighing of the pycnometer filled with distillate:

Weight of tare bottle at the time of

measurement

: T1= 171.9178 g

Pycnometer filled with distillate at 20.50 °C

: P2= 167.8438 g

Variation in the buoyancy of air

: dT = 171.9178 – 171.9160

= + 0.0018

Mass of the distillate at 20.5 °C

: Lt  = 167.8438 – (67.6695 + 0.0018)

= 100.1725 g

Apparent density of the distillate : ρ20.5 °C = 100.1725/101.8194 =

 0.983825 g/cm3

Calculation of alcoholic strength:

Refer to the table of apparent densities of hydro-alcoholic mixtures at different temperatures, as indicated above.

 

Part B Determination of the alcoholic strength by volume of a beverage by measuring the density of the distillate by electronic densimetry using a frequency oscillator

(Type I method)

(Resolution OENO 8/2000 – OIV/OENO 377/2009)

 

B.1.Principle

 

In the present method the distillate density is measured by electronic densimetry using a frequency oscillator. The principle consists of measuring the period of oscillation of a tube containing the sample undergoing electromagnetic stimulation. The density is thus calculated and is linked to the period of oscillation by the following formula:

(1)

ρ = density of the sample 

T = period of induced vibration

M = mass of empty tube

C = spring constant

V = volume of vibrating sample

This relationship is in the form ρ = A – B (2), so there is a linear relationship between the density and the period squared. The constants A and B are specific to each oscillator are estimated by measuring the period of fluids of known density.

B.2.Reagents and products

B.2.1Reference fluids

Two reference fluids are used to adjust the densimeter. The densities of the reference fluids should encompass the densities of the distillates to be analysed. A spread of greater than 0.01000 g/cm3 between the densities of the reference fluids is recommended.

The reference fluids for the determination of the ABV of vitivinicultural beverages by electronic densimetry are as follows:

  • dry air (unpolluted),
  • Type II water for analytical usage (ISO 3696 standard), or of equivalent purity,
  • hydro-alcoholic solutions of densities determined by another reference method, for which the uncertainty does not exceed 0.00005 g/cm3 at the temperature of 20.00 0.05°C,
  • solutions calibrated with traceability to the International System of Units (SI), with viscosities of less than 2 mm2/s, for which the uncertainty does not exceed 0.00005 g/cm3 at the temperature of 20.00 0.05°C.

B.2.2. Cleaning and drying products

Use products that ensure the perfectly clean and dried state of the measuring cell, according to the manufacturer’s indications. For example:

  • detergents, acids, etc.,
  • organic solvents: 96% vol. ethanol, pure acetone, etc.

B.3. Apparatus and equipment

B.3.1. Electronic densimeter with frequency oscillator

The electronic densimeter consists of the following elements:

  • a measuring cell consisting of a measuring tube and a temperature-controlled enclosure,
  • a system for setting up an oscillation tube and measuring the period of oscillation,
  • a digital display and possibly a calculator.

The densimeter is placed on a perfectly stable support isolated from all vibrations.

B.3.2. Temperature control of the measuring cell

Locate the measuring tube in the temperature-controlled enclosure. Temperature stability should be better than 0.02°C.

It is necessary to control the temperature of the measuring cell when the densimeter makes this possible because this strongly influences the determination results. The density of a hydro-alcoholic solution with an ABV of 10% vol. is 0.98471 g/cm3 at 20 °C and 0.98447 g/cm3 at 21 °C, equating to a spread of 0.00024 g/cm3.

The test temperature is set at 20 °C. The temperature is taken at the cell level, and done with a thermometer that has a resolution accurate to 0.01 °C and is calibrated to national standards. This should enable a temperature measurement with an uncertainty of better than 0.07°C.

B.3.3. Calibration of the apparatus

The apparatus should be calibrated before using it for the first time, then periodically or if the verification is not satisfactory. The objective is to use two reference fluids to calculate the constants A and B [see formula (2), B.1). To carry out the calibration in practice, refer to the user manual of the apparatus. In principle, this calibration is carried out with dry air (taking into account the atmospheric pressure) and very pure water (B.2.1).

B.3.4. Calibration verification

In order to verify the calibration, measure the density of the reference fluids.

Every day of use, a density check of the air is carried out. A difference between the theoretical density and observed density of more than 0.00008 g/cm3 may indicate that the tube is clogged. In that case, it should be cleaned. After cleaning, verify the air density again. If the verification is not conclusive, adjust the apparatus.

Check the density of the water; if the difference between the theoretical density and the density observed is greater than 0.00008 g/cm3, adjust the apparatus.

If verification of the cell temperature is difficult, it is possible to directly check the density of a hydro-alcoholic solution of comparable ABV to those of the distillates analysed.

B.3.5. Checks

When the difference between the theoretical density of the reference solution (known with an uncertainty of 0.00005 g/cm3) and the measured density is above 0.00008 g/cm3, the calibration of the apparatus should be checked.

B.4. Procedure

After obtaining a distillate, measure the density by densimetry and match to the ABV using the Tables.

Ensure the stability of the temperature of the measuring cell. The distillate in the densimeter cell should not contain air bubbles and should be homogeneous. If there is a lighting system available that makes it possible to verify the absence of bubbles, turn it off quickly after checking because the heat generated by the lamp can influence the measuring temperature.

For apparatus with a permanent lighting system, this statement is not applicable.

If the apparatus only provides the period, the density can be calculated from the A and B constants (see Annex I). If the apparatus does not provide the ABV directly, by knowing the density, obtain the ABV using the tables (Table I, OIV-MA-312-02).

B.5. Expression of results

B.5.1. Expression of results

B.5.1.1. Beverages with an ABV greater than or equal to 1.5% vol.

The alcoholic strength by volume of the beverage is obtained from the distillate. This is expressed as ‘% vol’.

If the temperature conditions are not respected, a correction should be made to express the temperature at 20 °C. The result is given to two decimal places.

B.5.1.2. Beverages with an ABV less than 1.5% vol.

Identical to B.5.1.1, dividing the alcoholic strength of the distillate (ABVD) by 2.

ABV = ABVD/2, % vol. to 2 d.p.

The validation parameters for beverages with a low alcohol content are given in Annex II

B.5.2. Comments

The volume introduced into the cell should be sufficient enough to avoid possible contamination caused from the previous sample. It is thus necessary to carry out at least two tests. If these do not provide results included in the repeatability limits, a third test is necessary. In general, the results from the last two tests are homogeneous and the first value can then be eliminated.

B.6. Precision

For samples with an ABV of greater than 4% vol., the validation data and precision results are given in Annex III.

epeatability (r) = 0.067 (% vol.)

Reproducibility (R) = 0.0454 + 0.0105 x ABV (% vol.)

For samples with an ABV of less than 4% vol. the validation data and precision results are given in Annex II

 

Part C: Determination of the alcoholic strength by volume of a beverage by measuring the density of the distillate by densimetry using a hydrostatic balance

(Type I Method)

(Resolution Oeno 24/2003 – OIV/OENO 377/2009)

C.1. Principle

The alcoholic strength by volume can be determined by densimetry using a hydrostatic balance following the Archimedes principle, by which any body immersed in a fluid experiences an upward force equal to the weight of the displaced fluid.

C.2. Reagents and products

 

C.2.1. Type II water for analytical usage (ISO 3696 standard), or of equivalent purity,

C.2.2. Floater-washing solution (sodium hydroxide, 30% m/v).

To prepare a 100 mL solution, weigh out 30 g of sodium hydroxide and fill to volume using 96% vol. ethanol.

C.3.           Common laboratory apparatus, including the following:

C.3.1.     Single-pan hydrostatic balance with 1 mg precision.

C.3.2.     Floater with at least 20 mL volume, specifically adapted for the balance, suspended by a thread with a diameter of less than or equal to 0.1 mm.

C.3.3.     Cylindrical test tube with level indicator.

The floater should be able to fit entirely within the test tube volume below the level indicator; only the hanging thread should break the surface of the liquid. The cylindrical test tube should have an inside diameter at least 6 mm greater than that of the floater.

C.3.4.     Thermometer (or temperature-measurement probe) with degree and 10th-of-a-degree graduations, from 10 °C to 40 °C, calibrated to ± 0.05 °C.

C.3.5.     Masses calibrated by an accredited body.

C.4.            Procedure

After each measurement, the floater and the test tube should be cleaned with distilled water, wiped with soft laboratory paper that does not lose its fibres and rinsed with solution whose density is to be determined. These measurements should be carried out once the apparatus has reached a stable level in order to limit alcohol loss through evaporation.

C.4.1.     Apparatus calibration

C.4.1.1.                       Balance calibration

While balances usually have internal calibration systems, hydrostatic balances should be calibrated with weights with traceability to the International System of Units (SI).

C.4.1.2 Floater calibration

Fill the cylindrical test tube up to the level indicator with water (C.2.1) whose temperature is between 15 °C and 25 °C, but preferably at 20 °C.

Plunge the floater and the thermometer into the liquid, shake, note down the density on the apparatus and, if necessary, adjust the reading in order for it to be equal to that of the water at the measurement temperature.

C.4.1.3. Control using a hydro-alcoholic solution

Fill the cylindrical test tube up to the level indicator with a known titre of hydro-alcoholic solution at a temperature of between 15°C and 25 °C, preferably at 20°C.

Plunge the floater and the thermometer into the liquid, shake and note down the density on the apparatus (or the alcoholic strength if possible). The established alcoholic strength should be equal to the previously determined alcoholic strength.

Note: This solution of known alcoholic strength can also replace water for floater calibration.

C.4.2.     Measurement of the density of the distillate (or alcoholic strength if possible)

Pour the test sample into the cylindrical test tube up to the level indicator.

Plunge the floater and the thermometer into the liquid, shake and note down the density on the apparatus (or the alcoholic strength if possible).

Note the temperature if the density is measured at t °C (ρt).

Correct ρt using a ρt density table of hydro-alcoholic mixtures (Table II in Annex I of the method OIV-MA-AS312-02 in the OIV Compendium of International Methods of Analysis).

C.4.3.     Cleaning of the floater and cylindrical test tube

Plunge the floater into the washing solution in the test tube.

Allow to soak for one hour while turning the floater regularly.

Rinse with tap water, then with distilled water.

Wipe with soft laboratory paper that does not lose its fibres.

Carry out these operations when the floater is used for the first time and then on a regular basis as necessary.

C.5.           Expression of results

C.5.1.     Beverages with an ABV greater than or equal to 1.5% vol.

Using the density ρ20, calculate the real alcoholic strength using the table indicating the alcoholic strength by volume (% vol.) at 20°C according to the density at 20°C of the hydro-alcoholic mixtures. This is the international table adopted by the International Organization of Legal Metrology in its Recommendation No. 22 (1973).

The values are expressed in % vol. to 2 d.p.

C.5.2.     Beverages with an ABV less than or equal to 1.5% vol.

Identical to C.5.1, dividing the alcoholic strength of the distillate (ABVD) by 2.

ABV = ABVD/2, % vol. to 2 d.p.

The validation parameters for beverages with a low alcohol content are given in Annex II.

C.6.           Precision

Repeatability (r)=  0.074 (% vol.)

Reproducibility (R)=  0.229 (% vol.)

 

Part D: Determination of the alcoholic strength by volume of a wine by measuring the density of the distillate by hydrometry or refractometry

(Type IV Method)

 

D.1.           Principle

The alcoholic strength may be determined by densimetry using an alcoholometer following the Archimedes principle. A weighted cylinder equipped with a graduated stem is more or less immersed into the distillate whose density is to be determined. The density of the liquid is read directly on the graduation of the stem at the level of the meniscus.

 

D.2.           Hydrometry

D.2.1.    Apparatus and materials

D.2.1.1.                      Alcoholometer

The alcoholometer should meet the specifications for Class I or Class II alcoholometers as defined in OIML (International Organization of Legal Metrology) International Recommendation 44 “Alcoholometers and alcohol hydrometers for use in alcoholometry”.

This apparatus should be calibrated with traceability to the International System of Units (SI).

D.2.1.2.                       Thermometer calibrated with traceability to the International System of Units (SI) with degree and 10th-of-a-degree graduations, from 0 °C to 40 °C, calibrated to ± 0.05 °C.

D.2.1.3.                       Measuring cylinder with dimensions that allow for the immersion of the thermometer and the alcoholometer without contact with the sides, held vertically.

D.2.2.     Procedure

Pour the distillate into the measuring cylinder, ensure that the cylinder is kept vertical, and insert the thermometer and alcoholometer. Stir and wait 1 minute to allow temperature equilibration of the measuring cylinder, the thermometer, the alcoholometer and the distillate before reading the thermometer. Remove the thermometer and, after 1 minute of rest, read the apparent alcoholic strength.

Take at least three readings from the bottom of the meniscus using a magnifying glass. Correct the apparent strength measured at t °C to account for the effect of the temperature using the Tables. The temperature of the liquid must differ very little from the room temperature (at most, by 5 °C).

 

D.3.           Refractometry

D.3.1.    Apparatus

Refractometer enabling the refractive indices in the range 1.330 to 1.346 to be measured.

Depending on the type of apparatus, measurements are taken:

either at 20°C, with a suitable instrument,

or at room temperature, t°C, with a thermometer enabling the temperature to be determined to within at least 0.05°C (a temperature correction table will be provided with the apparatus).

D.3.2.     Procedure

The refractive index of the wine distillate (5) is measured by following the procedure prescribed for the type of instrument used.

D.3.3.     Expression of results

Table IV in Chapter OIV-MA-AS312-02 is used to find the alcoholic strength corresponding to the refractive index at 20°C.

Note: Table IV gives the alcoholic strengths corresponding to the refractive indices for both pure hydro-alcoholic mixtures and for wine distillates. In the case of wine distillates, it takes into account the presence of impurities in the distillate (mainly higher alcohols). The presence of methanol lowers the refractive index and thus the alcoholic strength.

Note: To obtain the alcoholic strength from the density of the distillate, use Tables I, II and III in Chapter OIV-MA-AS312-02. These have been calculated from the international alcoholometric tables published in 1972 by the International Organization of Legal Metrology in its Recommendation No. 22 and adopted by the OIV.

 

7.      Annexes

 

Annex I

Formula for the calculation of alcoholic strength tables for mixtures of ethanol and water.

The density, , ρ expressed in kilograms per cubic metre (kg/m³) of a mixture of ethanol and water at the temperature t, expressed in degrees Celsius, is given by the following formula, according to the following:

the mass concentration ρ, expressed by a decimal number (*),

the temperature t, expressed in degrees Celsius (IPTS 68),

the numerical coefficients in the tables below.

The formula is valid for temperatures of between –20 °C and +40 °C.

(*) E.g. For a mass concentration of 12%, p = 0.12.

Numeric coefficients for the formula

 

Une image contenant table

Description générée automatiquement

 

Annex II Validation parameters relating to the measurement of the ABV of beverages with a low alcohol content

This document presents the results of the validation study for the method for beverages with a low alcohol content (update).

The study was carried out in accordance with documents OIV MA-F-AS1-08-FIDMET and MA-F-AS1-09-PROPER.

  1. Sample

Sample no.

1

2

3

4

5

6

Nature

Grape juice

Beverage obtained by dealcoholisation of wine

Beverage obtained by partial dealcoholisation of wine

Partially fermented grape juice

Cider

Wine-based beverage

Approximate ABV in %vol.

< 0.5

0.5

1.5

2.5

4.5

6.5

Table 1: Samples analysed for the validation

  1. Analyses

Each of the 12 samples received by the laboratories were analysed by simple distillation or by steam distillation according to the following two procedures:

OIV reference method with use of 200 mL and recovery of 200 mL of distillate,

Alternative method with use of 200 mL and recovery of 100 mL of distillate.

  1. Participating laboratories

 

19 laboratories from different countries took part:

Laboratório CVRVV

4050-501 Porto

Portugal

Laboratório de Análises da CVRA

7006-806 Évora

Portugal

Testing Laboratory CAFIA

603 00 BRNO

Czech Republic

Laboratório ASAE - LBPV

1649-038 Lisboa

Portugal

Agroscope - Site de Changins

1260 Nyon 1

Switzerland

Labo SCL de Bordeaux

33608 Pessac

France

Labo SCL de Montpellier

34196 Montpelllier

France

Laboratorio Arbitral Agroalimentario

28023 Madrid

Spain

Estación Enológica de Haro

26200 Haro La Rioja

Spain

Instituto dos Vinho do Douro do Porto

Porto 4050-253

Portugal

IVICAM

13700 Tomelloso, Ciudad Real

Spain

INCAVI

08720 Vilafranca del Penedès

Spain

ICQRF Laboratorio di Conegliano/Susegana

31058 SUSEGANA (TV)

Italy

ICQRF Laboratorio di Catania

95122 CATANIA

Italy

ICQRF Laboratorio di Modena

41100 Modena

Italy

ICQRF laboratorio di Perugia

06128 Perugia

Italy

ICQRF laboratorio di Salerno

84098 Salerno

Italy

ICQRF Laboratorio centrale di  Roma

00149 Rome

Italy

Laboratoires DUBERNET

11100 Narbonne

France

Table 2: Laboratories that took part in the validation

 

  1. Results

Sample No. 1

Sample No. 2

Sample No.°3

Sample No. 4

Sample No. 5

Sample No. 6

LAB

POSITION :

2

7

4

11

6

12

5

8

9

10

1

3

A

0.21

0.21

0.55

0.55

1.34

1.34

2.58

2.58

4.59

4.60

6.54

6.50

B

0.11

0.14

0.49

0.50

1.32

1.38

2.60

2.57

4.68

4.72

6.52

6.55

C

0.33

0.28

0.68

0.61

1.43

1.35

2.63

2.60

4.63

4.66

6.58

6.51

D

0.62

0.62

1.38

1.36

2.68

2.67

4.69

4.73

6.62

6.64

E

0.20

0.21

0.55

0.56

1.36

1.40

2.61

2.62

4.67

4.68

6.56

6.55

F

0.18

0.12

0.52

0.51

1.31

1.30

2.56

2.56

4.70

4.66

6.51

6.54

G

0.22

0.22

0.55

0.56

1.37

1.37

2.62

2.62

4.68

4.68

6.58

6.57

H

0.41

0.42

1.25

1.27

2.46

2.49

4.57

4.56

6.39

6.40

I

0.20

0.13

0.54

0.48

1.32

1.28

2.60

2.58

4.62

4.62

6.57

6.55

J

0.24

0.24

0.58

0.60

1.41

1.37

2.63

2.63

4.69

4.67

6.55

6.55

K

0.22

0.22

0.56

0.55

1.35

1.35

2.63

2.63

4.67

4.68

6.59

6.58

L

0.22

0.23

0.56

0.57

1.38

1.36

2.63

2.61

4.66

4.67

6.56

6.57

M

0.18

0.18

0.53

0.53

1.33

1.29

4.66

4.65

6.53

6.52

N

0.22

0.23

0.56

0.57

1.38

1.41

2.26

2.61

4.67

4.67

6.51

6.57

O

0.12

0.19

0.53

0.52

1.33

1.33

2.64

2.62

4.67

4.67

6.51

6.55

P

0.25

0.25

0.57

0.58

1.39

1.41

2.66

2.65

4.70

4.68

6.62

6.62

Q

0.22

0.20

0.55

0.59

1.34

1.33

2.61

2.63

4.65

4.63

6.52

6.54

R

0.21

0.21

0.55

0.52

1.29

1.28

2.52

2.55

4.62

4.56

6.50

6.53

S

0.18

0.17

0.41

0.42

1.38

1.37

2.61

2.58

4.63

4.58

6.51

6.48

Table 3: Results obtained for a 200 mL distillation with recovery volume of 200 mL

Results not presented were rejected in accordance with the Cochran (variance outliers) test with a 2.5% significance level (one-tailed test) and the Grubbs (outliers from the mean) test with significance levels of 2.5% (two-tailed test).

Note: The absent values have not been provided by the laboratory in question.

Sample No. 1

Sample No. 2

Sample No.°3

Sample No. 4

Sample No. 5

Sample No. 6

LAB

POSITION :

2

7

4

11

6

12

5

8

9

10

1

3

A

B

0.17

0.18

0.52

0.53

1.34

1.36

2.62

2.62

4.62

4.60

6.48

6.52

C

0.25

0.25

0.56

0.62

1.35

1.36

2.50

2.46

4.48

4.44

6.12

6.19

D

0.29

0.29

0.63

0.63

1.43

1.42

2.66

2.65

4.68

4.69

6.58

6.59

E

0.24

0.24

0.58

0.58

1.39

1.39

2.64

2.64

4.66

4.67

6.55

6.57

F

0.21

0.18

0.53

0.53

1.31

1.27

2.41

2.48

4.30

4.31

6.22

5.89

G

0.24

0.24

0.56

0.57

1.35

1.36

2.58

2.57

4.57

4.56

6.46

6.43

H

0.19

0.18

0.48

0.55

1.33

1.32

2.51

2.55

4.59

4.54

6.38

6.42

I

0.25

0.18

0.56

0.53

1.34

1.33

2.62

2.61

4.64

4.64

6.25

6.28

J

0.24

0.24

0.55

0.56

1.31

1.32

2.49

2.53

4.37

4.34

6.14

6.12

K

0.25

0.25

0.57

0.57

1.37

1.38

2.60

2.61

4.60

4.61

6.48

6.38

L

0.24

0.24

0.55

0.55

1.35

1.31

2.52

2.47

4.38

4.31

6.09

6.06

M

0.19

0.20

0.55

0.55

1.34

1.31

4.68

4.67

6.52

6.54

N

0.28

0.26

0.58

0.59

1.28

1.28

2.52

2.47

4.44

4.32

6.01

6.15

O

0.19

0.25

0.57

0.57

1.39

1.39

2.63

2.64

4.66

4.66

6.57

6.57

P

0.25

0.26

0.57

0.57

1.36

1.36

2.58

2.56

4.54

4.53

6.34

6.38

Q

0.24

0.24

0.57

0.57

1.38

1.38

2.63

2.62

4.66

4.67

6.56

6.56

R

0.23

0.23

0.54

0.55

1.32

1.30

2.54

2.56

4.56

4.52

6.40

6.35

S

0.27

0.26

0.55

0.57

1.34

1.34

2.46

2.43

4.53

4.51

6.36

6.36

Table 4: Results obtained for a 200 mL distillation with recovery volume of 100 mL

 

Results not presented were rejected in accordance with the Cochran (variance outliers) test with a 2.5% significance level (one-tailed test) and the Grubbs (outliers from the mean) test with significance levels of 2.5% (two-tailed test).

Note: The absent values have not been provided by the laboratory in question.

Sample 1

Sample 2

Sample 3

Sample 4

Sample 5

Sample 6

No. of laboratories selected

17

19

19

17

19

18

No. of repetitions

2

2

2

2

2

2

Minimum

0.11

0.41

1.25

2.46

4.56

6.48

Maximum

0.33

0.68

1.43

2.68

4.73

6.64

Overall average

0.20

0.54

1.35

2.60

4.65

6.55

Repeatability variance

0.00052

0.00033

0.00050

0.00019

0.00036

0.00047

Reproducibility variance

0.00211

0.00345

0.00190

0.00229

0.00181

0.00147

Inter-laboratory standard deviation

0.043

0.057

0.041

0.047

0.040

0.035

Repeatability standard deviation

0.02

0.02

0.02

0.01

0.02

0.02

r limit

0.06

0.05

0.06

0.04

0.05

0.061

Repeatability CV

11.1

3.3

1.7

0.5

0.4

0.3

Reproducibility standard deviation

0.046

0.059

0.044

0.048

0.043

0.038

R limit

0.130

0.166

0.123

0.135

0.120

0.109

Reproducibility CV

22.5

10.9

3.2

1.8

0.9

0.6

Horwitz RSDr

3.36

2.90

2.52

2.29

2.09

1.99

Horratr

3.3

1.1

0.7

0.2

0.2

0.2

Horwitz RSDR

5.10

4.39

3.82

3.46

3.17

3.01

HorratR

4.4

2.5

0.8

0.5

0.3

0.2

Table 5: Data obtained for a 200 mL distillate from a 200 mL sample

Sample 1

Sample 2

Sample 3

Sample 4

Sample 5

Sample 6

No. of laboratories selected

16

15

18

17

17

17

No. of repetitions

2

2

2

2

2

2

Minimum

0.17

0.52

1.27

2.41

4.30

6.01

Maximum

0.29

0.63

1.43

2.66

4.69

6.59

Overall average

0.24

0.56

1.35

2.56

4.55

6.38

Repeatability variance

0.00006

0.00003

0.00016

0.00050

0.00039

0.00135

Inter-laboratory standard deviation

0.03209

0.02496

0.03752

0.07013

0.12167

0.17621

Reproducibility variance

0.001

0.001

0.001

0.005

0.015

0.031

Repeatability standard deviation

0.01

0.01

0.01

0.02

0.02

0.04

r limit

0.02

0.02

0.04

0.06

0.06

0.104

Repeatability CV

3.2

1.0

0.9

0.9

0.4

0.6

Reproducibility standard deviation

0.033

0.025

0.039

0.072

0.122

0.178

R limit

0.092

0.071

0.109

0.203

0.347

0.504

Reproducibility CV

13.8

4.5

2.9

2.8

2.7

2.8

Horwitz RSDr

3.27

2.88

2.52

2.29

2.10

2.00

Horratr

1.0

0.4

0.4

0.4

0.2

0.3

Horwitz RSDR

4.96

4.36

3.82

3.47

3.18

3.03

HorratR

2.8

1.0

0.8

0.8

0.9

0.9

Table 6: Data obtained for a 100 mL distillate from a 200 mL sample

Annex III Validation parameters relating to the measurement of the ABV

by electronic densimetry (Part B)

  1. Inter-laboratory tests: precision and accuracy on additions 

1.1.  Samples

The samples used for this joint study are described in Table 7.

No.

Nature

Approximate ABV (% vol.)

C0

Cider (filtered through a membrane to remove CO2)

~5

V0

Filtered wine

~10

V1

Filtered wine then doped

~11

V2

Filtered wine then doped

~12

V3

Filtered wine then doped

~13

P0

Liqueur wine

~16

Table 7: Samples for the joint study

All samples were homogenised before filling the bottles to be sent to the participants. For wine, 40 litres were homogenised before sending and carrying out the additions.

For the additions, absolute ethanol was poured into a 5-L volumetric flask, then filled up to the line with filtered wine. This was repeated two times. The volumes of ethanol were 50, 100 and 150 mL respectively for the V1, V2 and V3 samples.

1.2.  Participating laboratories

The participating laboratories in the joint study are outlined in Table 8.

Laboratory

Postcode

City

ALKO Group LTD

FIN-00101

Helsinki

Bénédictine

76400

Fécamp

Casanis

18881

Gemenos

CIVC

51200

Epernay

Cointreau

49181

St Barthélémy d'Anjou

Courvoisier

16200

Jarnac

Hennessy

16100

Cognac

IDAC

44120

Vertou

Laboratoire Gendrot

33000

Bordeaux

Martell

16100

Cognac

Ricard

94320

Thiais

SOEC Martin Vialatte

51319

Epernay

Table 8: List of laboratories participating in the joint study

In order to not introduce a methodological bias, the results of the Station Viticole du Bureau National Interprofessionnel du Cognac (the joint-study organiser) are not taken into account.

1.2.1.      Analyses

The C0 and P0 products were distilled independently two times, and the V0, V1, V2 and V3 products three times. Three ABV tests were done for each distillate. The results are displayed in the results table.

1.2.2.      Results

The second test (out of the three carried out) was kept for the precision study (Table 9).

 Laboratory

C0

V0

V1

V2

V3

P0

6.020

9.500

10.390

11.290

12.100

17.080

1

5.970

9.470

10.380

11.260

12.150

17.080

9.450

10.340

11.260

12.150

6.040

9.500

10.990

11.270

12.210

17.050

2

6.040

9.500

10.390

11.280

12.210

17.050

9.510

10.400

11.290

12.200

5.960

9.460

10.350

11.280

12.170

17.190

3

5.910

9.460

10.360

11.280

12.150

17.200

9.450

10.340

11.260

12.170

6.020

9.470

10.310

11.250

12.160

16.940

4

6.020

9.450

10.350

11.250

12.120

17.070

9.450

10.330

11.210

12.130

5.950

9.350

10.250

11.300

12.050

17.000

5

5.950

9.430

10.250

11.300

12.050

17.000

9.430

10.250

11.300

12.050

6.016

9.513

10.370

11.275

12.222

17.120

6

6.031

9.513

10.336

11.266

12.222

17.194

9.505

10.386

11.275

12.220

5.730

9.350

10.230

11.440

12.080

17.010

7

5.730

9.430

10.220

11.090

12.030

16.920

9.460

10.220

11.080

11.930

5.990

9.400

10.340

11.160

12.110

17.080

8

6.000

9.440

10.320

11.150

12.090

17.110

9.440

10.360

11.210

12.090

6.031

9.508

10.428

11.289

12.180

17.089

9

6.019

9.478

10.406

11.293

12.215

17.084

9.509

10.411

11.297

12.215

6.030

9.500

10.380

11.250

12.150

17.130

10

6.020

9.510

10.380

11.250

12.150

17.100

9.510

10.380

11.250

12.160

6.020

9.480

10.400

11.260

12.150

17.040

11

6.000

9.470

10.390

11.260

12.140

17.000

9.490

10.370

11.240

12.160

Table 9: Results (second test per distillate) (% vol.)

1.2.3.      Repeatability and reproducibility calculations

The repeatability and reproducibility calculations were carried out in compliance with the standard NF X 06-041, September 1983, ISO 5725. Table 10 presents the standard deviation per cell (laboratory x sample).

Laboratory

C0

V0

V1

V2

V3

P0

1

0.0354

0.0252

0.0265

0.0173

0.0289

0.0000

2

0.0000

0.0058

0.3436

0.0100

0.0058

0.0000

3

0.0354

0.0058

0.0100

0.0115

0.0115

0.0071

4

0.0000

0.0115

0.0200

0.0231

0.0208

0.0919

5

0.0000

0.0462

0.0000

0.0000

0.0000

0.0000

6

0.0106

0.0046

0.0255

0.0052

0.0012

0.0523

7

0.0000

0.0569

0.0058

0.2050

0.0764

0.0636

8

0.0071

0.0231

0.0200

0.0321

0.0115

0.0212

9

0.0085

0.0176

0.0115

0.0040

0.0202

0.0035

10

0.0071

0.0058

0.0000

0.0000

0.0058

0.0212

11

0.0141

0.0100

0.0153

0.0115

0.0100

0.0283

Table 10: Dispersion table (standard deviation in % vol.)

Three cells presented strong dispersions (probability with Cochran test under 1%). These cells are represented in grey (Table 10). For laboratory 7 and the V3 product, the standard deviation of 0.0764 was maintained despite the Cochran test because it is on the same high level as that observed at the same laboratory for the V0 product.

An examination of the figures for each distillate (Table 9) led to the elimination of the following:

  • laboratory 2, product V1, value 10.990,
  • laboratory 7, product V2, value 11.440.

After eliminating these two values, the cell averages were calculated (laboratory x sample). The results are presented in Table 11.

Laboratory

C0

V0

V1

V2

V3

P0

1

5.9950

9.4733

10.3700

11.2700

12.1333

17.0800

2

6.0400

9.5033

10.3950

11.2800

12.2067

17.0500

3

5.9350

9.4567

10.3500

11.2733

12.1633

17.1950

4

6.0200

9.4567

10.3300

11.2367

12.1367

17.0050

5

5.9500

9.4033

10.2500

11.3000

12.0500

17.0000

6

6.0235

9.5103

10.3640

11.2720

12.2213

17.1570

7

5.7300

9.4133

10.2233

11.0850

12.0133

16.9650

8

5.9950

9.4267

10.3400

11.1733

12.0967

17.0950

9

6.0250

9.4983

10.4150

11.2930

12.2033

17.0865

10

6.0250

9.5067

10.3800

11.2500

12.1533

17.1150

11

6.0100

9.4800

10.3867

11.2533

12.1500

17.0200

Table 11: Table of averages (means in % vol.)

The figures given by laboratory 7 are generally low (Table 11). In the case of cider, the average for this laboratory is very far from the figures of the other laboratories (probability with the Dixon test under 1%). This laboratory's results for this product were eliminated.

Table 12 presents the calculated repeatability and reproducibility.

Sample

P

n

ABV

S2r

S2L

r

R

lC0

10

20

6.002

0.000298

0.001033

0.049

0.103

V0

11

33

9.466

0.000654

0.001255

0.072

0.124

V1

11

32

10.344

0.000255

0.003485

0.045

0.173

V2

11

32

11.249

0.000219

0.003113

0.042

0.163

V3

11

33

12.139

0.000722

0.003955

0.076

0.194

P0

11

22

17.070

0.001545

0.004154

0.111

0.214

Table 12: Repeatability and reproducibility calculations

Key:

p: number of laboratories retained

n: number of values retained

ABV: mean ABV (% vol.)

r: repeatability variance (% vol.)2

L: interlaboratory variance (% vol.)2

r: repeatability (% vol.)

R: reproducibility (% vol.)

Reproducibility increases with the sample's ABV (Figure 2). The increase in repeatability according to ABV is less noticeable and the overall repeatability was calculated according to the mean repeatability variance. As such, for the samples with an ABV of between 4 and 18% vol.,

Repeatability (r) = 0.067 (% vol.),

Reproducibility (R) = 0.0454 + 0.0105 x ABV.

1.2.4.      Accuracy with regard to additions carried out on wine

The regression line of alcoholic strength after addition according to the volume of ethanol added provides, for a volume of 0 mL, an estimation of the initial alcoholic strength of the product (Figure 3). This regression is carried out with mean values for each laboratory (Table 11).

Measurements carried out on initial products are not included in this estimation. This estimation was compared with the mean of the measurements taken on this product before additions; the intervals of relative confidence on these two estimations were calculated (Table 13).

LB

mean of measurements

UB

LB

estimation with measurements of products + additions

UB

9.440

9.466

9.492

9.392

9.450

9.508

Table 13: Additions to products

Key:

LB: lower bound of confidence interval at 95%

UB: upper bound of confidence level at 95%

The two confidence intervals cover a large overlapping centre. Thanks to the measurements on 'doped' samples, the alcoholic strength by volume of the initial product could be found.

1.2.5.      Conclusion of inter-laboratory tests

The repeatability and reproducibility indications by inter-laboratory tests provide the following equations, for products with ABVs of between 4% and 18% vol.:

 Repeatability (r) = 0.067 (% vol.),

 Reproducibility (R) = 0.0454 + 0.0105 x ABV (% vol.).

The Horwitz indicators, Hor and HoR, are low (Table 14). This therefore indicates good precision of the method in relation to the analyte measured.

Sample

C0

V0

V1

V2

V3

P0

n

20

33

32

32

33

22

p

10

11

11

11

11

11

ABV

6.0019

9.4662

10.3443

11.2492

12.1389

17.0699

r

0.0489

0.0724

0.0452

0.0419

0.0760

0.1113

sr

0.0173

0.0256

0.0160

0.0148

0.0269

0.0393

RSDr

0.2878

0.2702

0.1543

0.1316

0.2214

0.2303

RSDrH

2.0159

1.8822

1.8573

1.8340

1.8131

1.7224

Hor

0.1428

0.1436

0.0831

0.0718

0.1221

0.1337

R

0.1033

0.1237

0.1731

0.1634

0.1935

0.2136

sR

0.0365

0.0437

0.0612

0.0577

0.0684

0.0755

RSDR

0.6080

0.4616

0.5912

0.5131

0.5634

0.4423

RSDRH

3.0543

2.8519

2.8141

2.7788

2.7471

2.6097

HoR

0.1991

0.1619

0.2101

0.1847

0.2051

0.1695

 

Table 14: Summary table of method precision

Key:

n:

number of values retained

p:

number of laboratories retained

ABV:

mean ABV (% vol.)

r:

repeatability (% vol.)

sr:

repeatability standard deviation (% vol.)

RSDr:

repeatability coefficient of variation (sr x 100 / ABV) (%)

RSDrH:

Horwitz repeatability coefficient of variation (0.66 x RSDRH) (%)

Hor:

Horrat repeatability value (RSDr/RSDrH)

R:

reproducibility (% vol.)

sR:

reproducibility standard deviation (% vol.)

RSDR:

reproducibility coefficient of variation (sR x 100 / ABV) (%)

RSDRH:    Horwitz reproducibility coefficient of variation (2(1-0.5log(ABV))) (%)

HoR:         Horrat reproducibility value (RSDR/RSDRH)

The measurements carried out during inter-laboratory tests on wine with additions made it possible to find the value obtained before the addition. The values 9.45% and 9.47% vol. were found respectively.

 

Annex IV

Comparison of measurements carried out using a hydrostatic balance (Method C) with those obtained by electronic densimetry (Méthode B)

Using samples with alcoholic strengths between 4% vol. and 18% vol., the repeatability and reproducibility were measured using an inter-laboratory test. The alcoholic strength of the different samples as measured using a hydrostatic balance and using electronic densimetry were compared, including the repeatability and reproducibility values derived from the multi-year inter-comparison tests performed on a large scale.

  1. Samples: Wines with different densities and alcoholic strengths prepared monthly on an industrial scale, taken from a stock of bottles stored under normal conditions, and supplied anonymously to the laboratories.
  1. Laboratories: Laboratories participating in the monthly tests organised by Unione Italiana Vini (Verona, Italy) according to ISO 5725 (UNI 9225) regulations and the International Harmonized Protocol for the Proficiency Testing of Analytical Chemical Laboratories produced by the AOAC, ISO and IUPAC (J. AOAC Intern., 1993, 74/4), and ISO 43 and ILAC G13 guidelines. An annual report is provided by the above-mentioned organisation to all participants.
  1. Apparatus:
  • An electronic hydrostatic balance (with precision to 5 decimal places), equipped if possible with a data-processing device. 
  • An electronic densimeter, equipped if possible with an autosampler.
  1. Analyses

The measurement of the distillate was repeated twice.

  1. Results

Table 15 shows the results of the measurements obtained by the laboratories using a hydrostatic balance.

Table 16 shows the results obtained by the laboratories using an electronic densimeter.

  1. Evaluation of results

Interlaboratory reproducibility

A Horrat value of 1 usually indicates satisfactory interlaboratory reproducibility, whereas a value of more than 2 normally indicates unsatisfactory reproducibility, i.e. one that is too variable for analytical purposes or where the variation obtained is greater than that expected for the type of method employed.  Hor is also calculated and used to assess interlaboratory reproducibility, using the following approximation:

 

RSDr (Horwitz) = 0.66 RSDR (Horwitz) (this assumes the approximation that r= 0.66 R).

Table 17 shows the differences between the measurements obtained by laboratories using an electronic densimeter and those using a hydrostatic balance. Excluding the sample 2000/3, which has a very low alcohol strength and for which both techniques show poor reproducibility, good concordance is generally observed for the other samples.

  1. Precision parameters

Table 18 shows the overall averages of the precision parameters calculated from all monthly tests carried out between January 1999 and May 2001.

 In particular:

Repeatability (r) = 0.074 (% vol.) for the hydrostatic balance and

0.061 (% vol.) for electronic densimetry,

Reproducibility (R) =  0.229 (% vol.) for the hydrostatic balance and

0.174 (% vol.) for electronic densimetry.

Une image contenant table

Description générée automatiquement

Une image contenant table

Description générée automatiquement

Une image contenant table

Description générée automatiquement

 

Table 18: Precision parameters

MEAN

Hydrostatic balance

Electronic densimetry

 n1

441

557

 relative repeatability variance

0.309

0.267

 r

0.074

0.061

 sr

0.026

0.022

 relative reproducibility variance

2.948

2.150

 R

0.229

0.174

 sR

0.082

0.062

  1. Bibliography

Distillation:

  • HANAK, A., Chem. Zgt., 56, 1932, p. 984.
  • COLOMBIER, L., CLAIR, E., Ann. Fals. Fraudes, 29, 1936. p. 411.
  • POZZI-ESCOT, E., Ind. Agr. Aliment., 66, 1949, p. 119.
  • JAULMES, P., Analyse des vins, 1951, p. 49.
  • SCHNEYDER, J., Mitt. Klosterneuburg. Rebe und Wein., 10, 1960, p. 228.
  • SCHNEYDER, J., KASCHNITZ, L., Mitt. Klosterneuburg. Rebe und Wein., 15, 1965, p. 132.

Pycnometry:

  • JAULMES, P., Analyse des vins, 1951, p. 67.
  • JAULMES, P., Trav. Soc. Pharm. Montpellier., 12, 1952, p. 154.
  • JAULMES, P., Ann. Fals. Fraudes. 46, 1953, p. 84; 47, 1954, p. 191.
  • JAULMES, P., CORDIER, Mlle S., Trav. Soc. Pharm. Montpellier, 16, 1956, p. 115; 20, 1960, p. 137.
  • JAULMES, P., BRUN, S., Ann. Fals. Exp. Chim., 56, 1963. p. 129.

Hydrostatic balance:

  • Cabanis, M. T., Cassanas, G., Raffy, J. and Cabanis, J. C., 'Validation de la mesure du titre alcoométrique volumique', F.V. 1096, 1999.
  • Cabanis, M. T., Cassanas, G., Raffy, J. and Cabanis, J. C., 'Intérêt de la balance hydrostatique “nouvelle génération” pour la détermination du titre alcoométrique des vins et des boissons spiritueuses', Rev. Franç. Œnol., 177, July-August, 1999, pp. 28-31.
  • Versini, G. and Larcher, R., 'Comparison of wine density and alcoholic strength measurement by hydrostatic balance and electronic density–meter', Communication by the OIV "Methods of Analysis" Sub-Commission, Paris, 13-15 March 2002.
  • OIV Compendium of International Methods of Analysis of Wines and Musts E.g. Tables ??
  • International Office of Vine and Wine, Paris.
  • 'International Harmonized Protocol for the Proficiency Testing of Analytical Chemical Laboratories', J. AOAC Intern., 74/4, 1993.
  • ISO 5725 standards and ISO 43 guides.
  • OIV Resolution OENO 6/1999.
  • Horwitz, W., 'Protocol for the design, conduct and interpretation of method-performance studies', Pure and Applied Chemistry, 67/2, 1995, pp. 331-343.

Refractometry:

  • NEWTON, W., MURNO, F.L., Can. Chem. Met., 17, 1933, p. 119.
  • SAMPIETRO, C., INVERNIZZI, I., Ann. Chem. Appl., 30, 1940, p. 381.
  • FISCHL, P. F., Food Manufacture, 17, 1942, p. 198.
  • JAULMES, P., LAVAL, J. P., Trav. Soc. Pharm. Montpellier., 21, 1961, p. 21.
  • JAULMES, P., BRUN, S., LAVAL, J. P., Ann. Fals. Exp. Chim., 58, 1965, p. 304; Bull. Union National. OEnologues, 13, 1964, p. 17.
  • Alcoholic strength tables:
  • TABLES ALCOOMETRIQUES FRANCAISES, J.O. Républ. Française, 30 Dec 1884, P. 6895.
  • OIML, International Organisation of Legal Metrology, ‘International alcoholometric tables’, R22, 1973.
  • WINDISCH, K., based on LUNGE, G., BERL, E., Chem. techn. Untersuchungs Methoden, Berlin, 1924, 7th ed.; 4, 1893. p. 274.
  • OSBORNE, N. S., MCKELVY, E. C., BEARCE H. W., Bull. Bur. of Standards, Washington, 9, 1913, p. 328.
  • FROST, A. V., ‘Recherches dans le domaine du poids spécifique des mélanges d'alcool éthylique et d'eau’, Institute of Pure Chemical Reagents. U.S.S.R., No. 9, 1930, based on SPAEPEN, J.
  • HEIDE, C von der, MANDLEN, H., Z. Untersuch. Lebensm., 66, 1933, p. 338.
  • KOYALOVICS, B., 8th General Conference on Weights and Measures, Moscow, 1933.
  • FERTMANN, G. I., ‘Tables de renseignements pour le contrôle de la fabrication de l'alcool’, Pischerpoomizdat, Moscow, 1940.
  • REICHARD, O., Neue Alkohol u. Extract. Tafel 20°/20°, Verlag Hans Carl, Nürnberg, 1951.
  • JAULMES, P., MARIGNAN, R., Ann. Fals. Fraudes., 46, 1953, pp. 208 and 336.
  • SPAEPEN, J., Rev. de Métrologie, 1955, p. 411; Bull. belge de Métrologie, April No., 1955.
  • JAULMES, P., BRUN, S., Ann. Fals. Exp. Chim., 46, 1963, p. 143; 48, 1965, p. 58; 49, 1966, p. 35; 50, 1967, pp. 101-147; Trav. Soc. Pharm. Montpellier, 26, 1966, pp. 37 and 111.
  • JAULMES, P., MARIGNAN, R., Bull. O.I.V., 274, 1953, 28, 32.
  • JAULMES, P., BRUN, S., TEP, Y., Trav. Soc. Pharm., 28, 1968, p. 111.
  • KAWASAKI, T., MINOVA, Z., INAMATSU, T., ‘A new alcohometric specific gravity table’, National Research of Metrology, Tokyo, 1967.
  • TEP, Y., ‘Étude d'une table alcoométrique internationale’, Thèse Doc. Pharm, Montpellier, 1968.